MDC-60x cable


Communication protocolNone
Last spectrometerActive
Voltage protection

No voltage protection

Connector ALemo FGT.1M.308.XLMT
Connector BLemo FGT.1M.308.XLMT
Dust capsLemo BGF.1M.100.CAV or BGF.1M.100.XAV (note: this dust cap protects the connector on the cable)


The MDC-600 communication cables can be used to connect a PUCK-400(-RTK) and a spectrometer. Note that the cables are not equipped with an under- and overvoltage protection circuit and should therefore only be used in a setup where an MDC cable with power protection is used to power the PUCK. The MDC-600 cable connects power and ethernet from the PUCK to the spectrometer.

The Lemo FGT.1M.308.XLMT connector requires special tools to (dis)assemble. Please contact Medusa if the cable needs servicing.

Variants

MDC-600Can be used to connect the PUCK-400 to the spectrometer

Pin numbering

Lemo FGT.1M.308.XLMT
Note: Pin numbering as seen from the back side.

Pin description (Lemo connector)

#DescriptionWire color
1Power Red and blue
2Debugging onlyBrown / white
3Debugging onlyBlue / white
4Tx+Orange / white
5Tx-Orange
6Rx+Green / white
7Rx-

Green

8GroundBlack and brown

Notes

-

Datasheets