Skip to content
You're viewing an older version of this document. View the latest version
Created: 24 January 2023-Updated: 24 January 2023

LEDs

The detector has four LEDs embedded in the top cap. The orientation of the LEDs relative to the connector depends on the detector series. Some detectors are mounted with the fastening wings down such as the MS-2000, and some are mounted with the fastening wings up such as the MS-350 or the MS-1000. The position described below is with the connector oriented below the LEDs.

NameColorFunction
PWRGreenInput voltage is ok and in the right input range
RedInput voltage is too low
Red blinkingBlinking at 1 hz means that the input power could not be measured
BlueInput voltage is too high, possibly damaging the internal electronics
PurpleInput voltage has large fluctuations, please check the power supply
ETHGreenEthernet cable is connected
BlueThe device is connected to the internal through an internal 4g modem and simcard
SCNYellowA raw spectrum is collected.
GreenA raw spectrum is stabilized, analysed and stored.
AUXYellowA PTH reading is collected and stored
GreenA GPS reading is collected and stored
BlueA auxiliary reading is collected (possibly external sensor)

If the SCN and AUX LEDs blink in red this means that the internal storage is almost full. The threshold percentage for this signal as well as the action taken when the device is full can be set on the settings page on the webserver. The default action when all storage space is taken is to overwrite the oldest data.