MDC-80x cable

Communication protocolNone
Last spectrometerActive
Voltage protectionNo voltage protection
Connector ALemo FGG.1K.308/Lemo FGT.1M.308.XLMT
Connector BDepending on the type

The MDC-80x cables are used in the gSMS and RhoC5 systems. Note that not all cables in this series are interchangeable.

Please contact Medusa if the cable needs servicing.

Variants

MDC-800
(Deprecated)
Has a Lemo FGG.1K.308 on both sides. Usually supplied in combination with a solar panel and battery unit.
In gSMS systems this cable can be used for battery voltage monitoring
MDC-801
(Deprecated)
Has a Lemo FGG.1K.308 on one side. The other side has flying leads for power and an ethernet connector.
MDC-802
(Deprecated)
Used only to power the sensor via USB. Has a Lemo FGG.1K.308 on one side and a USB-A connector on the other side.
MDC-803Used only to power the sensor via power delivery USB-C. Has a Lemo FGG.1K.308 on one side and a USB-C connector on the other side. Use the pin-out of MDC-40x cables.

Pin description

MDC-800MDC-801MDC-802MD-803DescriptionWire color
1118GroundBlack or brown
2221PowerRed or blue
3n.c.n.c.n.c.Battery groundBrown/white
4n.c.n.c.n.c.Battery senseBlue/white
5Rx-n.c.n.c.Rx-Green
6Rx+n.c.n.c.Rx+Green/white
7Tx-n.c.n.c.Tx-Orange
8Tx+n.c.n.c.Tx+Orange/white

Notes

To monitor the battery voltage with a MDC-800 cable, the battery ground should be attached to the battery's minus terminal, and the battery sense should be attached to the positive terminal.